Smile, Smart, Speed

고객의 만족을 위하여 최선을 다하는 기업, 정문사이언스입니다.

분석장비

HomeHOME > 회사소개 >분석장비

이온화에너지 측정 시스템

모델명 : BIP-KV100
  • - Ionization Energy Measurement System

상품상세정보

Specifications

Measurement Principle

Photoelectron Yield Spectroscopy (PYS method)

Measurement WL range

9.54  4.0eV130  310nm

Min. measurement interval

0.01eV

Measurement resolution

10fA  100mA

Sample chamber

Atmosphere, nitrogen, vacuum compatible/VG65,NW25 port for vacuum pump

Irradiation light area

approx. 1×3mm ( by slit and vertical aperture )

WL resolution ( WL width )

0.08eV4.2nm) * at 256nm

Sample arrangement

Horizontal , Number of setting sample: 1 pc

Sample size

Max. 60 x 60mm, Min. 10 x 10mm, Thickness : within 4mm

Door switch

For safety of Internal applied voltage electrode and for applied voltage for PMT

Monochromator structure

Nitrogen purge ( light source and high order light cut-filters built in)

WL drive

Stepping motor, sine bar method ( GPIB control )

Focal length

100mm * for monochromator

Aperture ratio

F=3.5 * for monochromator

Mechancial WL range

 400nm * In case of the grating 2400 lines/mm

WL accuracy

±1nm * Setting from long WL *WL accuracy & repeatability

Tubing introduction

IN2 ( Swagelok 1/4), OUT2 ( Swagelok 3/8 ), Relief valve, Flow meter, Stop valve

PY measuring device

Micro ammeter ( Sub-Femtoamp Remote SourceMeter )

*PY=Photoelectron Yield

Irradiation intensity detector

PMT supplied with certified spectral sensitivity data for 130 ~ 310nm )

Software control function

System control, photoelectron measuring function,cal. of ionization energy (work function), DOS

Software save function

Data binary save and redisplay function, save in text

 

■ Background measurement and sample measurement can be continuously performed by the sample slide mechanism. Measurement is possible with the preset atmosphere (atmosphere, nitrogen, vacuum) maintained.

 

■ Ultrahigh sensitivity of 13 digits (10 fA ~ 100 mA) has been achieved with use of PYS method.

 

■ A port mounting a commercially available dry vacuum pump and composite molecular pump is provided

 

■ The optional Measurement Range Extension Unit enables the system to measure from 3.4 eV.

 

 

Measurement Data


Screen of Measurement Conditions

 

 

 

 

Ionization energy measurement data of TiO 2 particles [7.4 eV]

 

 

Measurement Items

① Background measurement:

The light to be irradiated on the sample is measured in advance with a

photomultiplier, and the number of photons is calculated.

②Sample measurement

Monochromatic light at each wavelength is irradiated to the sample while the

voltage is applied to the sample for the electrons to be released easily.

③ Ionization energy data processing

 Y ∝ (hν-I)n

Y= Released numbers of electron measured / numbers of photon irradiated

h Planck's constant

ν Frequency

I Threshold

n depends on density of electrons in the high level edge in the occupied state

of the sample.Only 2 or 3 can be selected in the parameter n.



 

 

Standard configurations

 

●Deuterium lamp 30W, (MgF2 window, natural air cooling )

●Power supply for D2 lamp

●Nitrogen purge type monochromator

●Grating:1200 grooves/mm blazed at 150nm ( MgF2 coating type )

●High order light cut filters and shutter

●Flow meter ( 1 ~ 10L/min)

●Sample chamber (atmosphere, nitrogen atmosphere, and vacuum compatible)

●Sample stage: 62 x 62mm (PTFE, detachable))

●Sample stage slide mechanism (operated outside of sample chamber, travel distance 65mm )

●Plate electrode up-down mechanism ( travel distance 6mm ))

●Standard sample ( Gold leaf Au)

●PMT for measuring irradiation light intensity and high voltage power supply (applied voltage -300V)

●Minute current measurement device for Photoelectron yield measurement

・Minimum range: 1pA Minimum range resolution: 10aA ( when the range is set at 1pA )

・Max. range : 100mA Max. range resolution: 1.0μA ( when range is set at 100mA )

・Noise level: within 0.4 fAp-p(4×10-16 A noise)

●Controller (GPIB control)

●GPIB-USB cable

●Software ( Windows 10 compatible, English )

●PC for controlling the system Note PC Windows 10 64bit

 

 

 

Option

 

●Measurement range expansion unit ( for 3.4 ~ 7.5eV )

・Grating 2400 grooves/mm blazed at 240nm ( MgF2 coating type )