광소자 측정 장치

모델명 : ART-25
  • - Optical Elements Measuring Device

상품상세정보

Specifications


Wavelength range

350 ~ 850nm( option : 220 2000nm

Measurement mode

Transmittance 0 100 (T)

Measurement precision

Within ±0.2

Measurement method

Back ground measurement method

Correction method

Double beam 2 system correction

Wavelength purity

Approx. 2nm

Wavelength accuracy

±0.1n

Irradiation beam size

Approx.5×5mm

polarization angle

0°45°90°( to be set by software )

Light source

Halogen lamp 150

integrating sphere

Φ100mm

Detector

Photomultiplier




The standard transmittance measurement range is from 350 to 850nm.

And optionally it can be extended to 220 ~ 2000nm.

Measurement repeatability within ± 0.2% is achieved with our proprietary double beam optical

system which cancels variations of light source and detector.

It also supports transmittance measurement of the prism as an option, and can also measure transmittance of any prism such as a single lens reflex pentaprism or a video camera BGR decomposition prism.

In addition to P polarization and S polarization measurement, direct measurement of 45° N polarized light is possible.



Interference filter measurement data (example) 



Transmittance data of polarization 0° P, polarization 45°, 

polarization 90° of the interference filter




 

At wavelengths where 3 data polarization 0°P, polarization

45°N, polarization 90 ° S overlap each other, 3 data always

overlaps at 1 point.


Standard Configuration

●Halogen lamo 150W ●Halogen lamp power supply ( for 150W) ●Halogen lamp holder ( for 150W) ●Light converging optical system ●Monochromator (M25T) ●Grating 1200 grooves /500nm blaze ●Wavelength driving mechanism ●Auto shutter mechanism ●Auto high order light cut filter mechanism ●High order light cut filters ●Double beam optical system ●Polarizer (Glan-Thompson) ●Polarizer auto-rotation mechanism ●Integrating sphereΦ100mm ●Photomultiplier ●Sample stage ●JQA filters for calibration ( 4 types, with case ) ●Amp system for high precision measurement ●Inter face unit ●Large sample chamber ●Frame ●PC ●Dedicated software ●Instruction manual

Optional Items

●D2 lamp unit ( lamp, power supply, lamp holder ) ●Xenon lamp unit ( lamp, power supply, lamp holder ) ●Two lamps light converging optical system ●Different type of gratings ●Different type of high order light cut filters ●Silicon photo diode ( for optional long wavelength expansion ) ●Integrating sphere moving mechanism ●Various type of sample stage 

Installation Environment

・Main unit:Approx.W2100×D900×H1750mm 
・Power:AC100V±10V 50/60Hz 10A