Lab single channel syringe pump
Lab single channel syringe pump main
functions:
1. Syringe inner diameter input function:
You can add a syringe from the list or enter the syringe ID directly.
2. The human-machine interface of the
micro-injection pump is friendly: the large screen LCD display, The digital
knob is used in conjunction with the membrane button for quick and easy
operation.
3. Working modes: perfusion mode,
extraction mode, extraction perfusion mode, continuous mode.
4. Memory function:
(1)The EEPROM saves the setup parameters
and does not need to be reset after power-on.
(2)When the operation is running in the
flow mode, the power is turned off. After the power is restored, the parameters
can be continuously run or stopped according to the set parameters.
5. Protection function: When the propulsion
mechanism of the syringe pump is blocked during the work, the syringe pump will
stop the work of the propulsion mechanism and emit a whistle alarm.
6. It can be connected to the host computer
by 485 communication bus and controlled by the background software.
7. External control interface: with access
to / output control.
8.Calibration function: More accurate flow
is obtained through the calibration procedure.
9. Syringe protection function: The syringe
can be prevented from being damaged by adjusting the position of the limit
block.
Lab single channel syringe
pump technical parameters:
Micro syringe pump working mode |
Perfusion, extraction |
Number of syringes |
1 |
Maximum stroke |
140mm |
Stroke resolution |
0.156μm |
Line speed range |
5μm/min-65mm/min (flow rate = line speed
× syringe area) |
Line speed adjustment resolution |
5μm/min |
Stroke control accuracy |
Error ≤ ± 0.5% (stroke ≥ 30% of the
maximum stroke) |
Rated linear thrust |
>90N |
Syringe selection |
Built-in main manufacturers, the main
model of the syringe for selection |
Syringe definition |
May directly enter the inner diameter of
the syringe |
Flow correction |
Get a more accurate volume through the
calibration procedure |
Operating parameter settings |
Dispensing liquid volume, injection time,
etc. |
Display parameter selection |
Fluid volume, flow rate, or linear
velocity |
Power-down memory |
After power-on, you can choose whether to
continue working according to the state before power off. |
Status signal output |
2-channel OC gate signal output for
indicating start/stop and direction status |
Control signal input |
2 channel start/stop control input end, 1
channel falling edge trigger signal control start and stop; 1 channel TTL
level signal control start and stop |
Communication Interface |
RS485 |
Dimensions |
280×210×140 (mm) |
weight |
3.69kg |
Power supply |
AC 90V-260V/15W |
Working environment relative humidity |
<80% |
Working environment temperature |
0℃- 40℃ |