Online close-loop control oxygen analyzer

모델명 :

상품상세정보

Online close-loop control oxygen analyzer has a friendly man-machine dialogue menu, which is intuitive and convenient to operate.It has a 320×240 dot matrix color LCD display, the display content is exquisite and clear. The imported zirconium dioxide sensor is with high measurement accuracy, a rapid response speed and long calibration cycle for features. The imported built-in sampling pump has long serving life and is reliable for working. 

Online close-loop control oxygen analyzer main features:

1. The accurate measurement among the range of ppm ~ % can be met by one-point calibration in the air, making its calibration simple and convenient;

2. The sampling pump can be turned off automatically under the condition of gas circuit fault with the functions of gas circuit blocking alarm and self protection, which could extend the serving life of sampling pump and sensor efficiently;

3. Power being supplied at a wide range makes it more suitable for different occasions;

4. No reference gas is needed, and the analyzer wouldn’t be affected by the oxygen concentration of working around atmosphere;

5. Adopting RS232 (default) /RS485 (optional) bidirectional communication, it could do one-way or bidirectional communication with the upper computer;

6. The oxygen concentration in the gas protection furnace can be controlled automatically without manual control, making the response immediate;

7. The control error made by the pressure and temperature of the protection air source can be compensated automatically, making the components in the system more stable;

8. With the PID parameters self-setting function, users only need to do the self-setting operation, and the analyzer could calculate the optimal PID controlling parameter according to the relation between controlling output and oxygen concentration sampling input to accomplish the accurate and automatic control to the oxygen concentration;

9. It can save the protection air source at maximum, decreasing the maintenance cost efficiently;

10. Flow measurement function, with instantaneous flow collection and cumulative flow display, calculation, storage (storage once every 10 minutes) and clear (menu clear) function;

11. The meter is mainly used for nitrogen flow control of SMT main road. When the oxygen concentration in the core temperature zone is too low and the proportional valve dynamically adjusts the flow rate, the meter will automatically output a standard flow rate, so that the oxygen concentration in the furnace is always within the set range ;

12. With proportional valve inlet and outlet pressure detection function;

13. Bypass protection function, when the proportional valve fails, the bypass solenoid valve can be automatically opened in time to keep the shielding gas into the atmosphere protection furnace without manual intervention, and the response is timely and rapid, reducing the failure of the proportional valve. Loss;   

14. The upper computer software is optional with the functions of curve display, data saving, parameters setting and so on.

Online close-loop control oxygen analyzer technical parameters:

Test principle

zirconium dioxide

Display mode

320×240 matrix dot color LCD 

Test Range

0 ~ 10/100/1000ppm/ 1.00% / 25.00% O2

Accuracy

0~1000ppm/1.00% / 10.00% / 25.00% ≤±1% FS

0~100ppm≤±2% FS

0~10ppm≤±5%FS

Resolution

0.1ppm

 Repeatability

±1%FS

Response time

T90≤60S

Analog output

4~20mA.DC(non-isolated output, loaded resistance lower than 500 Ohm)

0~10V.DC(non-isolated output, loaded resistance higher than 10k Ohm)

2 way programmable alarm output (Dry contact type, maximum load 220VAC/2A)               

Other interface

RS232 (default) or RS485(optional), Ethernet

Power supply

AC100 ~ 240V 50/60Hz, power consumption is less than 60VA

Environment humidity

80RH

Sample gas temperature

-10 ~ +60

Sample gas flow

2 ~ 2.5l/min

Sample gas pressure

micro-positive pressure, micro-negative pressure or normal pressure

Sampling mode

extraction way

Background gas

N2 and inactive gaseous mixture

Appearance size

180mm×266mm×305mm(H×W×D)

Install hole cut

182mm×231mm(H×W), In the case of embedded installation

Air source pressure

 0.4 ~ 0.8MPa(PV inlet)

Sampling circuit port

NPT 1/8 internal thread

Controlling circuit port

NPT3/8 internal thread (PV connector)

Flow range

 0 ~ 65m3/h(normal)

Flow statistics range

0~72m3/h (including cumulative flow function)

Protection valve flow

30~60m3/h (air source pressure 4~8bar)