Split type trace oxygen analyzer
Split type trace oxygen analyzer has a
friendly man-machine dialogue menu, which is intuitive and convenient to
operate. The analyzer adopts an imported zirconia oxygen sensor, which has the
characteristics of no power consumption, no consumption life, high measurement
accuracy, long calibration interval period, fast response speed and small zero
drift. it uses large-screen dot matrix LCD, which can display parameters such
as oxygen amount, date and time at the same time.
Split type trace oxygen analyzer main
features:
1) Any point calibration within the range
can be used to meet the measurement accuracy of the entire range without
multi-point calibration;
2) Chinese and English menus, users can
switch freely;
3) Restore factory settings to prevent
confusion of analyzer parameter settings due to incorrect operation;
4) Automatic data storage function, which
allows customers to freely access historical data;
5) The upper and lower limit alarm points
can be set arbitrarily within the full range;
6) Standard RS232 (default) or RS485
communication interface,it can achieve two-way communication with the computer
(optional).
Split type trace oxygen analyzer
technical parameters:
Measuring principle |
zirconium dioxide |
Display mode |
128×64 dot matrix LCD |
Range selectable |
0~10/100/1000ppm, 1.00%, 25.00% O2 |
Resolution |
0.1ppm |
Measurement accuracy |
0~1000ppm /1.00%/10.00% /25.00% ≤±1% FS 0~100ppm ≤±2% FS 0~10ppm ≤±5%FS |
Repeatability |
≤±1% |
Response time |
T90<30S |
Analog output |
4~20mA.DC (non-isolated output, load
resistance less than 500 ohms) 0-10V output (non-isolated output, load
resistance greater than 10K ohms) 1 way programmable dry contact passive
alarm output, contact maximum capacity 220VAC/2A |
Communication interface |
RS232 (default) or RS485 |
Working power |
AC 170 ~ 265V 50 / 60Hz, power
consumption is less than 10VA |
Ambient temperature |
-10 ~ +50℃ |
Ambient humidity |
<80% RH |
Sample gas temperature |
0 ~ 200 ℃(in the case of sampling pipe is stainless steel pipe) |
Sampling method |
plug-in (diffusion) |
Sample gas pressure |
86-106KPa, steady pressure
atmosphere |
Sample gas component |
no flammable gas, no corrosive gas,
the amount of dust is less than 1mg/Nm3 |
Size |
155mm × 155mm × 118mm (H × W × D) |
Hole size |
135mm × 135mm (H × W) |
Service life |
>24 months (under normal use
conditions) |
Air circuit interface |
NPT 1/8 internal thread |
Installation method |
embedded installation |