Trace oxygen analyzer

모델명 :

상품상세정보

Trace oxygen analyzer adopt an unique protection air circuit which can protect the sensor efficiently and prevent the sensor from being exposed in the air for too long to affect its serving life. The imported fuel-cell oxygen sensor is with long serving life, a rapid response speed, high measurement accuracy and long calibration cycle for features, etc.

Trace oxygen analyzer main features:

1. Only one random point needs to be calibrated among the whole measurement range, and the measured accuracy among the whole measurement range can be met without multi-points calibration;

2. Function of built-in temperature compensation could eliminate the influence from the changes of environment • temperature on the measured accuracy;

3. Function of storing data automatically makes looking through historical data at any time possible; 

4. Function of shifting Chinese-English menu makes the operation visual and easy;

5. Built-in imported extraction pump (optional).

Trace oxygen analyzer technical parameters:

Test principle

micro-fuel cell

Display mode

128*64 matrix dot LCD

Test Range

0 ~ 10/100/1000ppm,1.00%,25.00% O2

Resolution

0.1ppm

Accuracy

0~1000ppm /1.00%/10.00% /25.00% ≤±1% FS

0~100ppm ≤±2% FS

0~10ppm ≤±5%FS

Repeatability

≤±1%

Response time

T90≤60S

Analog output

4~20mA.DC(non-isolated output, loaded resistance lower than 500 Ohm)

0~10V.DC(non-isolated output, loaded resistance higher than 10k Ohm)

1 way programmable alarm output (Dry contact type, maximum load 220VAC/2A)

Communication Port

RS232 (default) or RS485

Power supply

AC170 ~ 264V 50/60Hz 

Temperature of the sample gas

0 ~ 50

Sampling mode

Sample gas pass-in or extraction

Sample gas flow

600 ~ 800ml/min

Working pressure(without pump)

0.05MPa≤entrance displayed pressure≤0.35MPa, steady pressure environment

Working pressure(with pump)

micro-positive pressure, micro-negative pressure or ordinary pressure

Background gas

HeH2CO2N2 and inactive gas

Appearance Size

177mm×251mm×343mm (H×W×D)

Install Hole Cut

179mm×213mm (H×W)

Serving life

24 months(under normal use)

Gas circuit port

NPT 1/8 internal thread

Install mode

frame mount