Smile, Smart, Speed
고객의 만족을 위하여 최선을 다하는 기업, 정문사이언스입니다.
고객의 만족을 위하여 최선을 다하는 기업, 정문사이언스입니다.
1200℃ single heating zone high vacuum CVD system with 3-channel float flowmeter to supply gas CY-O1200-50IT-3F-HV
1200℃ single heating zone high vacuum CVD
system consists of a single heating zone tube furnace, a three-channel
float flow meter and a high vacuum molecular pump set. The tube furnace is
controlled by the precision temperature control instrument for PID temperature
control. It can edit the 30-segment raising and falling temperature program and
has the function of overheating and short circuit protection. The flanges on
both sides of the furnace tube are equipped with a digital vacuum gauge and a
mechanical pressure gauge, which can be used to control the atmosphere in the
furnace tube. At the same time, the tube furnace is operated by high-definition
true color touch screen, which is easy to use. Even non-professionals can
master the use of the instrument after simple training, which can greatly
improve your experiment efficiency.
The system is equipped with a three-channel
float flowmeter, which can mix 1~3 channels of gas to accurately prepare the
mixed gas required for the experiment. Another pressure gauge is installed on
the front of the float flowmeter to monitor the gas pressure and the mixing
tank pressure.
This system is also equipped with a
precision molecular pump set, which can achieve a vacuum of 1.0E-3pa level,
which can provide users with a purer and more stringent experimental
environment than ordinary mechanical pumps.
1200℃ single heating zone high vacuum
CVD system applications:
It is suitable for high-temperature
sintering, metal annealing, quality inspection, etc. in universities, research
institutes, industrial and mining enterprises, especially for CVD experiments
that require a variety of gas mixing atmospheres.
1200℃ single heating zone high vacuum
CVD system technical parameters:
Vacuum tube furnace |
Model |
CY-O1200-50IT |
|
Tube material |
High purity quartz |
||
Tube diameter |
50mm (Optional 60mm/80mm/100mm) |
||
Tube length |
800mm |
||
Furnace chamber length |
440mm |
||
Heating zone length |
400mm |
||
Constant temperature zone |
200mm |
||
Operating temperature |
0~1100℃ |
||
Temperature control accuracy |
±1℃ |
||
Temperature control mode |
30 or 50 segment program temperature
control |
||
Display mode |
LCD |
||
Sealing method |
304 stainless steel vacuum flange |
||
Flange interface |
1/4" ferrule connector, KF16/25/40
joint |
||
Vacuum |
4.4E-3Pa |
||
Power supply |
AC:220V 50/60Hz |
||
Gas supply system |
Model |
CY-3F |
|
Gas channel |
3-channel |
||
Measuring unit |
Gas float flowmeter |
||
Measuring range |
A channel: 0~100ml/min
H2 gas |
Remarks: If you need other ranges
or gas types, you need to specify when ordering. According to the
customer's specific requirements, the flow meter of the corresponding gas
type and range can be selected. |
|
B channel: 16~160
ml/min N2 gas |
|||
C channel: 25~250
ml/min Ar gas |
|||
Measurement accuracy |
±2.0% |
||
Pipe pressure resistance |
3MPa |
||
Working pressure difference |
50~300KPa |
||
Connecting pipe |
304 stainless steel |
||
Control valve |
304 stainless steel needle valve |
||
Interface specification |
1/4" ferrule connector for gas
inlet and outlet |
||
Exhaust system |
Model |
CY-GZK103-A |
|
Molecular pump |
CY-600 |
||
Backing pump |
Rotary vane pump |
||
Pumping rate |
Molecular pump: 600L/S |
Comprehensive gas pumping performance:
the vacuum can reach 1.0E-3Pa in 30 minutes. |
|
Rotary vane pump: 1.1L/S |
|||
Pumping interface |
KF40 |
||
Exhaust interface |
KF16 |
||
Vacuum measurement |
Compound vacuum gauge: Resistance gauge +
ionization gauge |
||
Ultimate vacuum |
1.0E-5Pa |
||
Power supply |
AC:220V 50/60Hz |