Smile, Smart, Speed
고객의 만족을 위하여 최선을 다하는 기업, 정문사이언스입니다.
고객의 만족을 위하여 최선을 다하는 기업, 정문사이언스입니다.
Electrospinning special high precision syringe pump
Electrospinning special high precision
syringe pump can install different sizes of syringes (sample injectors),
and the syringe pump is suitable for liquid transmission in high precision and
small flow.The device is easy to use and simple to operate, and the parameter
setting, modification and viewing of the fluid control can be completed on the
main interface of the boot. When entering parameters, you can select the model
and inner diameter of the syringe directly from the list, or enter it manually.
Electrospinning special high precision
syringe pump main features:
1. Flow range: 0.001ul / min - 43.349ml /
min
2. Syringe inner diameter input function:
You can select the syringe from the list or directly input the inner diameter
of the syringe.
3. The micro-injection pump has a friendly
man-machine interface: a large-screen liquid crystal display, and a digital
knob for use with the membrane button, which is easy and quick to operate.
4. Working mode: Single push working mode.
5. Memory function:
(1) The EEPROM saves the setting parameters
and does not need to be reset after power-on.
(2) When the power is running in the flow
mode, the power is turned off. After the power is restored, the parameters can
be continuously run or stopped according to the set parameters.
6. Protection function: The pump will stall
and give an alarm when the drive structure of the pump is blocked.
7. The 485 communication bus can be
connected to the host computer and controlled by the background software.
8. External control interface: with
input/output control function.
9. Calibration function: More accurate flow
can be obtained through the calibration procedure.
10. Syringe protection function: The
syringe can be prevented from being damaged by adjusting the position of the
limit block.
Electrospinning special high precision
syringe pump specifications:
Working mode |
Perfusion |
Flow range |
0.001ul / min 一 43.349ml/min |
Number of syringes |
1 |
Maximum stroke |
140mm |
Stroke resolution |
0.156µm |
Line speed range |
5µm/min-65mm/min(flow rate = line
speed × syringe area) |
Line speed adjustment resolution |
5µm/min |
Stroke control accuracy |
Error≤±0.5% (stroke ≥ 30% of the
maximum stroke) |
Rated linear thrust |
>90N |
Syringe selection |
Built-in main manufacturer's main
model syringes for selection |
Syringe definition |
May directly enter the inner
diameter of the syringe |
Flow correction |
Get a more accurate volume through
the calibration procedure |
Operating parameter settings |
Dispensing liquid volume,
injection time, etc. |
Display parameter selection |
Fluid volume, flow rate, or linear
velocity |
Power-down memory |
After power-on, you can choose
whether to continue working according to the state before power off. |
Status signal output |
2-channel OC gate signal output for
indicating start/stop and direction status |
Control signal input |
2 channel start/stop control input end, 1
channel falling edge trigger signal control start and stop; 1 channel
TTL level signal control start and stop |
Communication Interface |
RS485 |
Dimensions |
280×210×140(mm) |
weight |
4.5kg |
Power supply |
AC 200V-260V/35W |
Working environment relative humidity |
<80% |
Working environment temperature |
0℃- 40℃ |