Smile, Smart, Speed
고객의 만족을 위하여 최선을 다하는 기업, 정문사이언스입니다.
고객의 만족을 위하여 최선을 다하는 기업, 정문사이언스입니다.
High precision 10-channel micro laboratory syringe pump
High precision 10-channel micro
laboratory syringe pump main features:
1. Syringe inner diameter input function:
You can add a syringe from the list or enter the syringe ID directly.
2. The human-machine interface of the
micro-injection pump is friendly: the large screen LCD display, The digital
knob is used in conjunction with the membrane button for quick and easy
operation.
3. Four working modes: perfusion mode,
extraction mode, extraction perfusion mode, continuous mode.
4. Memory function:
(1)The EEPROM saves the setup parameters
and does not need to be reset after power-on.
(2)When the operation is running in the
flow mode, the power is turned off. After the power is restored, the parameters
can be continuously run or stopped according to the set parameters.
5. Protection function: When the propulsion
mechanism of the syringe pump is blocked during the work, the syringe pump will
stop the work of the propulsion mechanism and emit a whistle alarm.
6. It can be connected to the host computer
by 485 communication bus and controlled by the background software.
7. External control interface: with access
to / output control.
Calibration function: More accurate flow is
obtained through the calibration procedure.
Syringe protection function: The syringe
can be prevented from being damaged by adjusting the position of the limit
block.
High precision 10-channel micro
laboratory syringe pump technical parameters:
Working mode |
perfusion |
Flow range |
0.831ul / min a 21.675ml/min |
Number of syringes |
10 |
Maximum stroke |
140mm |
Stroke resolution |
0.156μm |
Line speed range |
5μm/min-130mm/min (flow rate = line speed
× syringe area) |
Line speed adjustment resolution |
5μm/min |
Stroke control accuracy |
Error≤±0.5% (stroke ≥ 30% of the maximum
stroke) |
Rated linear thrust |
>180N |
Syringe selection |
Built-in main manufacturer’s main model
syringes for selection |
Syringe definition |
You may directly enter the inner diameter
of the syringe |
Flow correction |
Get a more accurate volume through the
calibration procedure |
Operating parameter settings |
Dispensing liquid volume, injection time,
etc. |
Display parameter selection |
Fluid volume, flow rate, or linear
velocity |
Power-down memory |
After power-on, you can choose whether to
continue working according to the state before power off. |
Status signal output |
2-channel OC gate signal output for
indicating start/stop and direction status |
Control signal input |
2 channel start/stop control input, 1
channel falling edge trigger signal control start and stop; 1 channel
TTL level signal control start and stop |
Communication Interface |
RS485 |
Dimensions |
280×210×140(mm) |
weight |
4.5kg |
Power supply |
AC 200V-260V/35W |
Working environment relative humidity |
<80% |
Working temperature |
0℃- 40℃ |