Flue oxygen analyzer
Flue oxygen analyzer has a variety of
fault information prompting functions. Self-contained fault detection
function can improve troubleshooting efficiency and self-protection
ability. PID automatic temperature control circuit, fast and accurate heating,
can eliminate the influence of external environment on temperature
control. With its own fault detection function, it can improve the
efficiency of troubleshooting and self-protection ability.
Flue oxygen analyzer main features:
1.High-bright OLED active light display,
intuitive display;
2.The new zirconia sensor has extremely
small drift, fast response and high precision.
3. Add coating on the pipe wall of the
sensor contacting the flue gas to reduce the corrosion of the sensor caused by
the corrosive gas and extend the service life;
4.The contact part of the oxygen probe and
the flue gas adopts a new high-temperature alloy steel, and the oxygen probe
can be directly inserted into the furnace chamber below 650℃;
5.Two-point calibration can meet the
measurement accuracy of the entire range;
6.The probe and meter are IP54 protection
class (except he probe front end);
7.Users can freely set the 4-20mA (default)
or 0-10V (optional) output corresponding to the upper and lower limit value, so
that users can get higher precision analog output;
8.Standard RS232 (default) or RS485
(optional) communication interface for two-way communication with a computer or
other digital communication device.
Flue oxygen analyzer technical
parameters:
Measuring principle |
zirconia |
Display mode |
128×64 dot matrix OLED |
Measuring range |
0.01% ~ 21.00% O2 |
Measurement accuracy |
<±1.0%.FS |
Repeatability |
<±0.5%.FS |
Stability |
<±1%.FS (24 hours) |
Resolution |
0.01% |
Response time |
T90 ≤ 5S |
Analog output |
4~20mA.DC (default) or 0~10V.DC
(non-isolated output, 4 ~ 20mA load resistance ≤ 500 ohm, 0 ~ 10V load
resistance > 10K ohm) 2-way programmable dry contact passive
alarm output, contact maximum capacity 220VAC/2A |
Other interfaces |
RS232 (default) or RS485 |
Calibration method |
online calibration |
Power supply |
AC220V±10% 50/60Hz, ≤100W |
Working temperature |
Meter : -10 ~ +50℃ Probe:0 ~ 650℃ |
Background gas |
burning fumes (except for strong
corrosion and reducing gases) |
Hole size |
Φ70mm~Φ90mm |
Connection method |
Meter and probe connection cable
(standard 3 meters) |
Probe length |
Standard length is 1110mm (other lengths
can be customized) |
Service life |
>60 months (under normal use
conditions) |
Installation method |
Probe: plug-in type, meter: wall-mounted |