Smile, Smart, Speed
고객의 만족을 위하여 최선을 다하는 기업, 정문사이언스입니다.
고객의 만족을 위하여 최선을 다하는 기업, 정문사이언스입니다.
Infrared Carbon Monoxide Analyzer
Infrared carbon monoxide analyzer is
designed based on the principle that the gas has a selective absorption of
infrared light. It has the characteristics of small size, fast response, high
accuracy and easy to carry. It can continuously measure the concentration of CO
gas in indoor and outdoor ambient air.
Infrared carbon monoxide analyzer main
features:
1. Friendly communication menu makes the
operation visual and simple;
2. Long calibration interval cycle, high
accuracy and reliable stability;
3.It is of good selectivity to the tested
gas;
4.The imported electrochemistry sensor is
with relatively rare error, a rapid response speed, high accuracy and
long calibration cycle for features;
5.The compensation system for temperature
can eliminate the influence from environment temperature;
6.With the standard RS232(default)or RS485 communication port carried
by it,the analyzer can do bidirectional communication
with the computer;
Infrared carbon monoxide analyzer
technical parameters:
Test principle |
Infrared principle |
Display mode |
128×64 matrix dot LCD |
Test range |
0 ~ 1000ppm/10.00% CO |
Accuracy |
≤±1.2%FS |
Repeatability |
≤±1% |
Error |
<±1.2%FS |
Response time |
T90≤60S |
Analog output |
4-20mA.DC(non-isolated output, loaded
resistance lower than 500 Ohm). 0-10V.DC(non-isolated output,
loaded resistance higher than 10k Ohm). 1 way programmable alarm output (Dry
contact type, maximum load 220VAC/2A) . |
Communication port |
RS232(default)or RS485 |
Power supply |
AC170 ~ 264V 50/60Hz |
Environment temperature |
-10℃ ~
+40℃ |
Environment humidity |
<80%RH |
Sampling mode |
sample gas pass-in |
Sample gas temperature |
0 ~ 50℃ |
Sample gas flow |
400 ~ 800ml/min |
Sample gas pressure |
0.05Mpa≤entrance displayed
pressure≤0.35Mpa,steady pressure environment |
Appearance Size |
133mm×268mm×300mm(H×W×D) |
Install Hole Cut |
135mm×228mm(H×W) |
Serving life |
>24 months (under normal use) |
Gas circuit port |
NPT 1/8 internal thread |
Install mode |
frame mount |