Laboratory embedded oxygen monitor
Laboratory embedded oxygen monitor
main features:
1) Modular design, the user can choose the
altitude measurement module;
2) Compared with electrochemical and
ultrasonic, it has a long service life and fast response;
3) The sensor is not powered, does not
consume, has a long service life and is easy to maintain;
4) no reference gas is required,
independent of the operating environment oxygen concentration;
5) High-precision temperature automatic
compensation system to eliminate the influence of ambient temperature;
6) The measurement range is 0 ~ 40.0% O2 wide.
Laboratory embedded oxygen monitor
technical parameters:
Measuring principle |
ion current |
Display mode |
128×64 dot matrix LCD |
Measuring range |
0 ~ 40.0% O2 |
Measurement accuracy |
<±1%FS |
Repeatability |
<±1% |
Response time |
T90 ≤ 30S |
Analog output |
4-20mA.DC (non-isolated output, load
resistance less than 500 ohms) 1 way programmable dry contact passive
alarm output, contact maximum capacity 220VAC/2A |
Power supply |
18 ~ 36V DC, <5W |
Sample gas temperature |
-10 ~ +50℃ |
Ambient humidity |
<80% RH |
Sampling method |
free diffusion |
Sample gas component |
no flammable gas, no corrosive gas, the
amount of dust is less than 1mg/Nm3 |
Service life |
>5 years (under normal use
conditions) |
Installation method |
The probe is suction mounted The display header is embedded |
Optional parameters |
|
Atmospheric pressure |
30 ~ 120KPa |
Ambient temperature |
-40℃~ +85℃ |
Ambient humidity |
0 ~ 100% RH |