Multi-component gas analyzer
Multi-component gas analyzer has a built-in
original imported mass flow meter, the flow display is intuitive and
convenient, and the service life is long. The three measured media all have
temperature compensation function to reduce the influence of ambient
temperature on measurement accuracy.The analyzer has two display modes, Chinese
and English, which users can switch freely. The sensor has an automatic
protection function without manual intervention.
Multi-component gas analyzer main features:
1.Large-screen touch screen display, single
screen displays three kinds of media concentration values, flow values and
other related parameters;
2.Historical data storage function, users
can query historical data and historical curves locally at any time;
3.The user can choose the altitude module
to avoid the influence of atmospheric pressure changes on the measurement
accuracy;
4.H2O, CO users can choose not to use the
range sensor;
5.Each measuring medium corresponds to one
analog and alarm output;
6.Built-in alarm buzzer, redundant alarm
points for three media;
7. If the zirconia sensor is not energized,
it does not be consumed, has a long life, and is easy to maintain;
8.Imported film capacitive sensor, fast
response and high sensitivity;
9.No reference gas is required, and is not
affected by the working environment oxygen concentration;
10.Very good selectivity for the gas to be
measured.
Multi-component gas
analyzer technical parameters:
Display mode |
640 × 480 dot matrix LCD touch screen |
Measuring range |
oxygen concentration: 10.00~ 99.99% O2 Carbon monoxide concentration: 0~ 10ppm
CO Dew point: -80.0~ 20.0 °C (basic
type)
-100~ +20.0°C (extended type) |
Test accuracy |
oxygen concentration: ≤ ± 1% FS Carbon monoxide concentration: ≤ ± 1% FS Dew point: ≤±1.2%FS |
Repeatability |
≤±1% |
Response time |
oxygen concentration: T90 ≤ 15S Carbon monoxide: T90 ≤ 30S Dew point: T 90 ≤ 50S |
Analog output |
Three-way 4-20mA.DC (non-isolated
output, load resistance less than 500 ohms), corresponding to three test
media Three-way programmable dry contact
passive alarm output (contact maximum capacity 220VAC/2A), corresponding to
three test media |
Communication interface |
RS232 (default) or RS485 |
Sample gas flow |
oxygen / carbon monoxide channel 0.4 ~
0.6L / min Dew point channel 1.5 ~ 2L / min |
Sample gas pressure |
inlet pressure 0.1 ~ 0.5MPa, outlet
pressure is atmospheric pressure |
Flow display |
electronic flow meter |
Sampling method |
straight-through type |
Service life |
oxygen sensor > 5 years (under normal
use conditions) Carbon monoxide sensor > 3 years
(under normal use conditions) Dew point sensor > 5 years
(under normal use conditions) |
Size |
176mm × 350mm × 356mm (H × W × D) |
Hole Size |
176mm × 307mm (H × W) |
Power supply |
AV170 ~ 264V 50/60Hz, power consumption
is less than 20W |
Air circuit interface |
NPT 1/8 internal thread |
Installation method |
embedded |