Smile, Smart, Speed
고객의 만족을 위하여 최선을 다하는 기업, 정문사이언스입니다.
고객의 만족을 위하여 최선을 다하는 기업, 정문사이언스입니다.
Thermal conductive hydrogen analyzer
Thermal conductive hydrogen analyzer has a
friendly communication menu that makes the operation visual and
simple, The imported thermal conductive sensor or the micro-fuel cell
principle has high accuracy and long serving life.It is with long calibration
interval cycle, high accuracy and reliable stability for features.
Thermal conductive hydrogen analyzer
main features:
1.The built-in temperature compensation
sensor could reduce the error from temperature change on the measured accuracy;
2.With the thermal conduction principle,
the sensor’s serving life won’t be consumed when the power is off;
3.Function of storing 5000 history data
allows users to look through the local data at any time;
4.The installation mode of frame mount
makes it simple and convenient to install;
5.With power supplied at a wide range, the
influence from mains fluctuation on the analyzer can be eliminated;
6.Thermal conduction or electrochemcial
principle available.
Thermal conductive hydrogen analyzer
technical parameters:
Test principle |
thermal conductive principle/micro-fuel
cell |
Display mode |
128×64 matrix dot LCD |
Test range |
0 ~ 30/50/99.99% H2(thermal
conductive principle) 0 ~ 10/100/2000ppm H2(eletrochemical
principle) |
Resolution |
0.01%/1ppm |
Accuracy |
thermal conductive principle: ≤±1%FS micro-fuel cell principle: ≤±1.5%FS |
Repeatability |
≤±1% |
Response time |
T90≤60S |
Analog output |
4-20mA.DC(non-isolated output, loaded
resistance lower than 500 Ohm) 0-10V.DC(non-isolated output, loaded
resistance higher than 10k Ohm) 1 way programmable alarm output (Dry
contact type, maximum load 220VAC/2A) |
Communication port |
RS232(default)or RS485 |
Power supply |
AC170 ~ 265V 50/60Hz |
Sample gas temperature |
-10 ~ +50℃ |
Sampling mode |
sample gas pass-in |
Sample gas flow |
400 ~ 800ml/min |
Sample gas pressure |
0.05Mpa≤entrance displayed
pressure≤0.35Mpa, steady pressure environment |
Background gas |
various mixed gas |
Appearance Size |
155mm×155mm×290mm(H×W×D) |
Install Hole Cut |
135mm×135mm(H×W) |
Serving life |
thermal conductive type>36 months, micro-fuel cell type>24 months(under normal use) |
Gas circuit port |
NPT 1/8 internal thread |
Install mode |
frame mount |