Smile, Smart, Speed

고객의 만족을 위하여 최선을 다하는 기업, 정문사이언스입니다.

HomeHOME > 회사소개 >

Bench Top Vacuum Ultraviolet Spectrophotometer

상품상세정보

Specification

Light Source

Deuterium lamp 300w

Wavelength range

120 ~ 300nm

Measurement reproducibility

±0.3%T  *1

Wavelength reproducibility

Within 0.1nm

Sample Size

Minimum 3 x 3 mm ~ Maximum 12 x 12 mm

Sample thickness

Minimum 1mm ~ Maximum 3mm  *2

Measurement method

Semi double beam

Detector

PMT with a salicylic acid soda window

Setting of display

Scale (Y axis, X axis), Trace(moving cursor)

 

Overlay of the spectrum, Delete of spectrum display

Data Processing

Smoothing, Arithmetic, 1st ~ 3rd Derivatives,

 

Peak processing (Peak Find Peak Height, Peak Area, Data Dump)r

*1 Under the following conditions: wavelength 120~300nm, Scanning speed 60nm/min, 100% flatness

*2 Specifications of the holder shape and sample numbers to be placed can be customized as option.


■ Transmission measurement in the range of 120 to 300 nm (vacuum ultraviolet region)

■ Nitrogen purge type monochromator enables high resolution and high stability

■ Ideal for evaluation of special films and special substrates

 

Measurement Data


T% data of thick synthetic quartz SiO



 

 

Green Synthetic quartz (t-40mm)

Blue Synthetic quartz (t-1mm )

Synthetic quartz (SiO2) with thickness of 1mm and 40mm was measured.  

The difference in transmittance due to the difference in thickness can be 

accurately measured by our unique optical system.



T% of calcium fluoride (CaF2)




 


Calcium fluoride (CaF2) used for evaluating optical crystals and
 resists was measured. 

The wavelength range for F2 laser and ArF laser can be measured with good 

reproducibility. 


T% of films for food packaging

 

Green Polyethylene

Pink Polyvinyl chloride

Blue Polychlorinated vinylidene

 

A film with a thickness of about 10μm was measured. Thin samples such as polymers can be easily measured. 

The system can measure continuously from the vacuum ultraviolet region to the ultraviolet region, which is difficult to 

measure with a general-purpose ultraviolet-visible spectrophotometer.


High resolution measurement of air (O)


 

Air was measured at 0.1 nm resolution. The fine structure in the Schumann-Runge

absorption band of oxygen in the air has been clearly measured. 






Standard configurations

● Deuterium lamp 30W

● Power supply for Deuterium lamp

● KV-200(nitrogen purge) monochromator

● Sample compartment (for nitrogen purge and vacuum )

● Transmission measurement sample holder

● PMT(photomultiplier tube) with a salicylic acid soda window

● Power supply for PMT

● Controller (wavelength drive, filter switching )

● Software (compatible with Windows 7)

● PC for controlling the system

 Instruction Manual

 

Transmission measurement sample holder


 

 

Software

Measurement conditions setting window 



 


 

Measurement data window



Option


● Vacuum pump

● Various type of sample holders

● Various gratings

● Salicylic acid soda window