Smile, Smart, Speed
고객의 만족을 위하여 최선을 다하는 기업, 정문사이언스입니다.
고객의 만족을 위하여 최선을 다하는 기업, 정문사이언스입니다.
1200℃ High vacuum nickle base alloy tube furnace
Technical parameters of high vacuum
nickle base alloy tube furnace:
Working Temperature |
1150℃ |
No. of Heating Zones |
Three |
Length of Each Heating Zone |
300-400mm |
Length of Uniform Temperature Zone |
500 mm (Minimum) |
Material of vacuum tight working tube |
Inconel 600 or equivalent Nickle base
alloy suitable for above mentioned Temperature. Water cooling jackets on both
ends of the working tube to be provided. |
Inside dia. of working tube |
100 mm (Minimum) |
Wall thickness of working tube |
5 mm (Minimum) |
Temperature uniformity |
± 5℃(In Uniform
Temperature Zone) |
Temperature Control |
By 03 Thyristors and 03 independent
Programmable PID controllers (One for each heating zone), Minimum 8 Segments
or other equivalent temperature controllers. Note: The temperature controllers must be
selectable for all types of thermocouple. |
Safety Controller |
A safety controller must be installed in
the furnace to shut off the furnace in case of temperature shooting due
failure of temperature control |
Thermocouples |
03 K type thermocouples for controlling/
measuring the temperature of 03 heating zones and 01 K type thermocouple for
safety controller, all housed in SS-304 sleeves. Each thermocouple to be
fixed in the center of each heating zone while thermocouple for safety
controller to be fixed near the thermocouple of central heating zone. The
thermocouples should be installed in the furnace in such a way that they
should touch working tube but should not be damaged during splitting of the
furnace. |
Furnace Body & Structure |
Main structure: MS Channels Outer Body: Hot part made of MS, powder
coated (in furnace) with hammer gray, while Cold Part (Control panel and
other) made of MS powder coated (in furnace) with green or blue paint. Material of Insulation: Fiber Ceramic
Board and Blanket (Bricks made of heavy Ceramic must not be used). |
Vacuum System |
High vacuum to be generated by rotary
vane pump and oil diffusion pump incorporated in a vacuum station. The valves
in the vacuum station are to be operated manually. Two stage rotary vane pump: pumping rate:
5L/s; Ultimate vacuum:1x 10-3 mbar. Oil Diffusion Pump: pumping rate: 600L/s;
Ultimate vacuum: 1x 10-6 mbar. Vacuum Gauge & Controller: Vacuum
gauge fitted in the working tube should be able to measure the vacuum during
heating operation of the furnace. Measuring range: 1x 10-9 mbar. |
Electrical System |
Heating: 380V, 3 phases, 50Hz, Approx.
45KVA Vacuum: 220V, single phase, 50Hz |